Neumüller / EMV / Medizintechnik

Was ändert sich mit der neuen Medizin-/EMV-Norm?

2. Februar 2017, 16:24 Uhr | Nicole Wörner
Diesen Artikel anhören

Fortsetzung des Artikels von Teil 1

Was sind die konkreten Anforderungen?

Port

Basic Standard for Test Method

EN 60601-1-2:2007

IEC 60601-1-2: 2014

Professional Healthcare Facility Environment

Home Healthcare Environment

Enclosure Port

IEC 61000-4-2

6 kV (Contact discharge)

8 kV (Air discharge)

8 kV (Contact discharge)

15 kV (Air discharge)

8 kV (Contact discharge)

15 kV (Air discharge)

IEC 61000-4-3

80 to 2500 MHz /

10 V/m for LIFE-SUPPORTING

80 to 2700 MHz /

3 V/m

80 to 2700 MHz / 10 V/m

80 to 2500 MHz /

3 V/m for not LIFE-SUPPORTING

Frequency: up to 5785MHz Level: up to 28V/m

Frequency: up to 5785MHz Level: up to 28V/m

IEC 61000-4-8

50 or 60 Hz

3 A/m

50 or 60 Hz

30 A/m

50 or 60 Hz

30 A/m

AC input power port

IEC 61000-4-4

2 kV

2 kV / 100kHz

2 kV / 100kHz

IEC 61000-4-5

1 kV line to line 2 kV line to earth

1 kV line to line 2 kV line to earth

1 kV line to line 2 kV line to earth

IEC 61000-4-6

0.15 to 80 MHz / 10 V for LIFE-SUPPORTING

0.15 to 80 MHz / 3 V

0.15 to 80 MHz / 3 V

0.15 to 80 MHz / 3 V for not LIFE-SUPPORTING

6 V in ISM bands

6V in ISM bands & amateur radio bands

IEC 61000-4-11

<5 % UT for 0,5 cycle

<5 % UT for 0,5 cycle

<0 % UT for 0,5 cycle

at step: 45°

at step: 45°

40 % UT for 5 cycles

0 % UT for 1 cycle

0 % UT for 1 cycle

70 % UT for 25 cycles

70 % UT for 25/30 cycles

70 % UT for 25/30 cycles

<5 % UT for 5 s

<5 % UT for 250/300 cycles

<5 % UT for 250/300 cycles

DC input power port

IEC 61000-4-4

2 kV

2 kV / 100kHz

2 kV / 100kHz

IEC 61000-4-5

N/A

1 kV line to line

2 kV line to earth

1 kV line to line

2 kV line to earth

IEC 61000-4-6

0.15 to 80 MHz / 10 V for LIFE-SUPPORTING

0.15 80 MHz / 3 V

0.15 to 80 MHz / 3 V

0.15 to 80 MHz / 3 V for not LIFE-SUPPORTING

6V in ISM bands

6V in ISM bands & amateur radio bands

ISO 7637-2

N/A

N/A

ISO 7637-2:2011

Signal input/output parts port

IEC 61000-4-4

1 kV / 5KHz

1 kV / 100kHz

1 kV / 100 kHz

IEC 61000-4-5

N/A

2kV line to earth for outdoor

2kV line to earth for outdoor

IEC 61000-4-6

0.15 to 80 MHz / 10 V for LIFE-SUPPORTING

0.15 to 80 MHz / 3 V

0.15 to 80 MHz / 3 V

0.15 to 80 MHz / 3 V for not LIFE-SUPPORTING

6V in ISM bands

6V in ISM bands & amateur

Patient coupling port

IEC 61000-4-2

N/A

8 kV (Contact discharge)

15 kV (Air discharge)

8 kV (Contact discharge)

15 kV (Air discharge)

IEC 61000-4-6

N/A

0.15 to 80 MHz / 3 V

0.15 to 80 MHz / 3 V

N/A

6V in ISM bands

6V in ISM bands & amateur radio bands